Instruction Manual
4-35
IM WX102-01E
1
2
3
4
5
6
7
8
9
10
11
Index
DARWIN Hardware Configurator
4.16  Printing Settings (DR Only)
The following print setting can be made for DRs:
•  Recording Mode (Trend or Logging)
•  Trend Printing interval (Fixed or Automatic)
•  Chart speed
•  Chart speed 2
•  Header
•  Title
•  List Format
To open the Printing related setting screen (see below), click the Printing tab, or select 
Settings > SET (Regular) Settings > and either Recording Mode, Chart Speed, Chart 
Speed 2, Header, Title, or List Format, from the menubar. 
Title settings
Choose list 
printing items
Header settings
Recording
interval
Chartspeed 2 
settings
Chart speed settings
Printing tab
Recording
modes
Recording Mode Settings
The following types of recording mode can be selected. 
•  LOGGING (LOGGING MODE) :  measurement values will be printed out as digital 
values.
•  TREND (ANALOG TREND MODE) : 
 measurement values will be recorded as analog 
trends (dot recording) and printed out as digital 
values.
•  Selecting the recording interval
  This setting can only be done for the TREND mode.
  FIX : 
Recording takes place at intervals equal to the measurement period (scan interval). 
However, if the measuring period is 0.5 or 1 second, the recording interval is xed 
to 2 seconds. The measured data during the excess period is ignored.
  AUTO :   recording takes place at intervals automatically decided by measurement 
interval and chart speed.