User guide

7-176
IM WX101-01E
Setting Constants
Constants are used for computation equation on models with computation function (/M).
Up to 60 constants (K01 to K60) can be set (Thirty constants (K01 to K30) for the
standalone type).
The number of significant digits is 5 excluding the decimal point. If an exponent is used,
the mantissa and exponent must consist of 5 digits and 2 digits, respectively.
Congurable ranges:
–1.0000E+35 to –1.0000E-35
0
1.0000E-35 to 1.0000E+35
Constants tab
Selecting Events
Event/Action and Match Time settings are possible for the DA100, if the (/M1) Math
option is being used, or if a pulse module is attached.
For DA100 and DC100, Group settings are possible if the Math option is being used.
Group settings
Match time settings
Action settings
Edge/Level settings
Event settings
Remote, Relay, Timer, and Match Time number settings
Event
Select any event from among the following:
NONE — releases the event/action setting.
REMOTE
This menu is displayed only with the DI/DO modules that are recognized as system
modules.
When a remote control signal is applied, the action is executed.
There are 12 control signals available. Select any control sign
al from these numbers
for event setting. Remote control signal terminal numbers are applicable for this
selection.
7.9 Setting DARWIN