User guide

7-157
IM WX101-01E
1
2
3
4
5
6
7
8
9
Index
Setting the Recorder
7.8 Setting MV1000/MV2000
Security
Settings Description
OFF Disables POP before SMTP.
POPbeforeSMTP Enables POP before SMTP.
Address 1, Address 2
Enter the e-mail address. Multiple e-mail addresses can be entered in the box of one
recipient. When entering multiple addresses, delimit each address with a space. Up
to 150 characters can be entered.
Sender
Enter the sender e-mail address. You can enter up to 64 characters.
POP3 Settings
If you need to use POP before SMTP, specify the POP3 server.
POP3 Server name
Enter the host name or IP address of the POP3 server.
Port number
Unless specied otherwise, set the number to the default value. The default value is
110.
Login name
Enter the POP3 server login name.
Password
Enter the POP3 server login password. You can enter up to 32 characters.
Send delay [second]
Enter the wait time from POP3 server authentication until transmission. Set a value in
the range of 0 to 10 (seconds).
Login method
To send the POP3 server login password without encryption, set POP3 Login to
[PLAIN]. To send the password with encryption, set POP3 Login to [APOP].
Alarm
Specify the settings for sending e-mail when alarms occur.
Recipient1 and Recipient2
Set the e-mail recipients. For Recipient1 and Recipient2, select [ON] to send e-mail
or [OFF] to not send e-mail.
Active alarms
Sends an e-mail when an alarm occurs. You can select [ON] (send e-mail) or [OFF] (not
send e-mail) for alarms 1 to 4.
Include INST
Select [ON] to attach instantaneous value data when the alarm occurred.
Include source URL
Select [ON] to attach the source URL. Attach the URL when the Web server is
enabled.
Subject
Enter the subject of the e-mail using up to 32 alphanumeric characters. The default
setting is Alarm_summary.
Header1, Header2
Enter header 1 and header 2 using up to 64 characters.