User guide
7-84
IM WX101-01E
• Set Calibration (Release numbers 4 and later)
To use calibration management, select [ON].
Notification
You can specify how many days before the calibration due date you want to display
the calibration notication screen. You can set the number of days to a value between
1 and 10.
Renotification
You can specify the period at which to display the calibration notication screen. The
calibration notication screen will continue to appear until calibration is completed.
• Signature (Only on DXs with the /AS1 advanced security option)
Process Type
Set the process type to choose what kind of collection of measured data can be
signed.
Setting Description
Continuous You can sign each individual measured data file.
Batch You can sign a collection of all the measured data files from the start to stop
of a recording. However, you can only sign a file from the DX when the file
covers the measured data of an entire recording, from start to stop.
Sign from Recorder
Set the signature privilege range for DX key operations.
Setting Description
Off You cannot sign files from the DX.
Signature1 You can sign files from the DX using the Signature1 privileges.
Signature1+2 You can sign files from the DX using the Signature1 and Signature2
privileges.
Signature1+2+3 You can sign files from the DX using the Signature1, Signature2, and
Signature3 privileges.
Memory Stop at Signature
You can congure this setting so that a screen for making a signature (historical trend
display) appears when recording stops (memory stop).
Setting Description
On The historical trend display appears automatically at memory stop.
Off The display does not change at memory stop.
Note
You cannot set [Memory stop at signature] when:
• [Sign from recorder] is set to [Off].
Even when [Memory stop at signature] is set to [On], the historical trend display will not appear
when:
• The process type is [Batch] and the measured data is divided into multiple files.
• A user without signature privileges performs memory stop.
• The multi batch function (/BT2 option) is being used and batch overview mode is enabled.