IM WX1-08E
Error Messages during Acquisition
Code Description Cause/Corrective Actions
W[631] Data Lack Data was unexpectedly lost. Reduce the number of acquired data points or 
connected instruments, or lengthen the scan interval.
E[673] Cannot open communication Communication failed to open
Check whether the power to the connected measuring instruments with which 
you wish to open communications is ON, and whether the cables are connected 
correctly. Also check the following. 
• For Ethernet
 Is the address correct, is the TCP/IP protocol installed in Windows, and is the 
Ethernet card installed correctly? 
• For RS-232 and RS-422A
  Are the baud rate, port numbers (COM1–COM9), and addresses (RS422) 
correct? Is the PC serial port valid? Is the cable type correct?
E[674] Communication error Communication Errors
Same as E[673]. 
E[675] Communication time out Communication timeout
Same as E[673]. 
W[880] Do not specify communication 
port
Specify the communication parameters. 
W[881] Current connecting recorder 
configuration is mismatch!
The recorder is:
Change the parameters for the main unit.
W[882] Cannot login recorder Recorder login error
W[882] Cannot Get recorder 
information
Failed to get settings from the recorder. Check whether the communication status 
and connected instruments matches those specified in the software. If they do not 
match, perform automatic model determination again. 
3.3 Error Messages and Corrective Actions