Details of Functions
Automatic Model Determination
If you select a recorder (µR10000/µR20000) on the software and perform automatic
model determination, the recorder model, type, number of measurement channels, and
number of computation channels are applied to the Recorder Sett
ing tab.
Model and Number of Channels
Model Type No of meas ch No of calc ch
µR10000/µR20000/SR10000 Pen 1 8
1
µR10000/µR20000/SR10000 Pen 2 8
1
µR10000/µR20000/SR10000 Pen 3 8
1
µR10000/µR20000/SR10000 Pen 4 8
1
µR10000/µR20000/SR10000 Dot 6 12
1
µR20000 Dot 12 24
µR20000 Dot 18 24
µR20000 Dot 24 24
1
The SR10000 does not have MATH channels.
Scan Interval
A scan interval from 0.5 to 3600 seconds can be selected for each of the sixteen
recorders.
Note
When connecting to DAQLOGGER and acquiring data from the recorder, if Gate µR’s scan
interval is longer than that of DAQLOGGER, DAQLOGGER logs the same data as is within the
Gate µR scan interval. Therefore, it is recommended that Gate µR’s scan interval be set to a
value smaller than DAQLOGGER’s scan interval.
Communication Retry Operation
You can turn communication retry ON and OFF, and set the retry interval.
When Turning Communication Retry OFF
After detecting that communications were broken, it is not reconnected.
When Turning Communication Retry ON
After detecting that communications were broken, the software attempts to reconnect
after the retry interval has elapsed. Even if the reconnection resulted in an error,
the software waits again for the retry interval to elapse, then attempts another
reconnection. This process repeats until a valid connection is opened.
A
communication retry interval of 30 to 3600 seconds can be entered.
Retries are also performed on instruments with which a communication error occurred
during the rst communication.
At the point that communication is restored, the
decimal place, alarm values, and other necessary information are retrieved from the
μR, and data is acquired.
Note
This software considers communications to have been cut in the following circumstances.
If the software requests data from the recorder but no response is received within ten
seconds.
If a TCP/IP connection is detected or if login fails when attempting to connect to the recorder
via Ethernet.
Port Settings
The following port settings can be entered using the software.
Monitor server port
The TCP/IP port number used for communication with DAQLOGGER or Remote
Monitor.
3.2 Details of Functions