User Manual

ConnectingthePowerCord(ifthepowersupplyvoltagesuffixcodeis-2)
When Using an AC Adapter
Precautions to Be Taken While Connecting the Power Supply
Makesuretofollowthewarningsbelowwhenconnectingthepowersupply.Toprevent
electricshockanddamagetotheMV,observethefollowingwarnings.
WARNING
• Topreventelectricshock,makesurethatthepowersourceisturnedOFF.
• OnlyusethepowercordthatOmegaprovidesforusewiththeMV.
• ConfirmthatthepowersourcevoltagematchesthespecificationsoftheAC
adapter before connecting the power cord.
• WhenyoudonotplanonusingtheMVforawhile,removethepowercordfrom
the AC outlet.
• OnlyuseaOmegaACadapter.
• DonotplaceobjectsontopoftheACadapterorpowercord,andkeepthem
away from heat sources.
• Whenremovingtheplugfromthepoweroutlet,donotpullonthecord.Pull
from the plug. If the power cord is damaged, contact the dealer from whom you
purchased the cord.
Use a power supply that meets the following conditions:
Item Specifications
Rated supply voltage 100 to 240 VAC
Operating supply voltage range 90 to 264 VAC
Rated supply frequency 50/60Hz
Power supply frequency range 48to62Hz
Maximum power consumption 45 VA (100 V), 60 VA (240 V)
Connection Procedure
1.
CheckthatthepowerswitchisOFF.
2.
Connect the power cord and the protective ground cord to the power terminal.
Use round crimp-on lugs (designed for 4-mm screws) with insulation sleeves. The
appropriate tightening torque for the screws is 1.4 to 1.5 N/m.
Power cord
Functional ground terminal
Protective grounding cord
Note
Do not wire the protective grounding
cord to the functional ground terminal.
3.
Attach the power terminal cover (transparent), and fasten it with screws.
Connecting a Power Cable
27
M-4661