User Manual

Connecting a Power Cable
ConnectingthePowerCord(ifthepowersupplyvoltagesuffixcodeis-1)
Precautions to Be Taken While Connecting the Power Supply
Makesuretofollowthewarningsbelowwhenconnectingthepowersupply.Toprevent
electricshockanddamagetotheMV,observethefollowingwarnings.
WARNING
• Beforeconnectingthepowercord,ensurethatthesourcevoltagematchesthe
rated supply voltage of the MV and that it is within the maximum rated voltage
range of the provided power cord.
• ConfirmthatthepowerswitchisOFFbeforeconnectingthepowercord.
• Topreventelectricshockorfire,besuretousethepowercordsuppliedby
Omega.
• Makesuretoperformprotectivegroundingtopreventelectricshock.Connect
thepowercordofthedesktoptypetoathree-prongpoweroutletwitha
protective ground terminal.
• Donotuseanextensioncordwithoutprotectiveground.Otherwise,the
protection function will be compromised.
Use a power supply that meets the following conditions:
Item Condition
Rated supply voltage 100 to 240 VAC
Allowable power supply voltage range 90 to 132, 180 to 264 VAC
Rated power supply frequency 50/60Hz
Allowable power supply frequency range 50/60Hz±2%
Maximum power consumption 45 VA (100V) and 60 VA (240 V)
Note
Do not use a supply voltage in the range 132 to 180 VAC, as this may have adverse effects on
measurement accuracy.
Connection Procedure
1.
CheckthatthepowerswitchisOFF.
2.
Connect the power cord plug to the power connector on the rear panel. (Use the
powercordthatcomeswiththepackage.)
3.
Checkthatthepoweroutletmeetstheconditionsgiveninthetableaboveandthat
the supply voltage is within the maximum rated voltage range of the power cord
thatcomeswiththepackage.Then,connecttheotherendofthepowercordto
thepoweroutlet.Youmustuseathree-prongACoutletwithaprotectiveground.
26
M-4661