User guide

4-29
Commands
M-4662
1
2
3
4
5
6
7
App
Index
XJ SetsanRJC
Tousetheinternalcompensationcircuit
Syntax
XJ p1,p2<terminator>
p1
Measurement channel number
p2
Reference junction compensation
designation (INTERNAL)
Query
XJ[ p1]?
Example Set the channel 001 RJC to the internal
compensation circuit.
XJ001,INTERNAL
TouseanexternalRJC
Syntax
XJ p1,p2,p3<terminator>
p1
Measurement channel number
p2
Reference junction compensation
designation (EXTERNAL)
p3
External RJC value (–20000 to 20000)
Query
XJ[ p1]?
Example Set the channel 002 RJC to external and set the
compensation value to 0 μV.
XJ002,EXTERNAL,0
Description • Set parameter p1 according to the table in
section 4.3.
• The unit of parameter p3 is μV.
WU Setsenvironmentsettings
Setup items
GENERAL,BATCH,DISPLAY,MESSAGE,INPU
T, ALARM,SECURITY,MEDIA,MATH,REPORT,
SERVICEPORT,DECIMALPOINT,POP3
Generalenvironmentsettings
Syntax
WU p1,p2,p3<terminator>
p1
Setting type (GENERAL)
p2
Tag/channel number selection
TAG
CHANNEL
Channel number
p3
Language
ENGLISH
JAPANESE
CHINESE
GERMAN
FRENCH
KOREAN
Example Select tag display and set the language to
English.
WUGENERAL,TAG,ENGLISH
Tosetthebatchfunction
Syntax
WU p1,p2,p3,p4<terminator>
p1
Setting type (BATCH)
p2
Batch function ON/OFF
p3
Number of lot number digits (OFF, 4, 6, 8)
p4
Auto increment ON/OFF
Description Parameters p3 and p4 are valid only when p2 is
ON.
Tosetthedisplay
Syntax
WU p1,p2<terminator>
p1
Setting type (DISPLAY)
p2
Trend type (T-Y)
p3
Partial expansion ON/OFF
p4
Trend update rate switching ON/OFF
Tosetmessagesettings
Syntax
WU p1,p2,p3,p4<terminator>
p1
Setting type (MESSAGE)
p2
Method of writing messages from keys
COMMON
Applied to all display groups
SEPARATE
Applied to a specified display
group
p3
Power failure message ON/OFF
p4
Change message ON/OFF
Tosettheinput
Syntax
WU p1,p2<terminator>
p1
Setting type (INPUT)
p2
Detection of values exceeding the scale
FREE
When the measurement range is
exceeded
OVER
When ±105% of the scale is
exceeded
Tosetalarmsettings
Syntax
WU p1,p2<terminator>
p1
Setting type (ALARM)
p2
Alarm suppression function ON/OFF
Tosetsecuritysettings
Syntax
WU p1,p2,p3<terminator>
p1
Setting type (SECURITY)
p2
Keys
OFF
Disable security
KEYLOCK
Lock the keys
LOGIN
Enable the login function
p3
Communication
OFF
Disable security
LOGIN
Enable the login function
Setmediasettings
Syntax
WU p1,p2,p3<terminator>
p1
Setting type (MEDIA)
p2
Auto save function ON/OFF
p3
Media FIFO ON/OFF
Example Use media FIFO.
WUMEDIA,ON,ON
Description Parameter p3 is valid only when p2 is ON.
4.6BasicSettingCommands