Manual
4-2
M-4660
• EventDataRecordingConditions
PressMENUandthenselect>Menutab>Datasave>Eventdata.
Settings
• Memory>Data Kind
Setting Description
Display Records display data.
E+D Records display data and event data. You cannot select this setting if the
trend update interval switching function (Trend rate switching) is set to on
On (see section 6.3 for details)
Event Records event data.
• Memorysample>On/Off
Select On for the target channels.
• Dispdata>Trend/Storage interval
Set the display trend/storage interval. For the trend/storage intervals that can only be
set on high-speed input models, 5 s/div and 10 s/div, 1 division (div) is equal to 40
dots.
You can only set trend/storage intervals that are longer than the scan interval you set
in Basic Setting Mode.
• Dispdata>Save interval (when recording display data)
Set the size of recorded data les. Recorded data is divided into les of the size
specied here. The values that you can set here vary depending on the Trend/Storage
interval setting.
Trend/Storage
interval
1
5s
2
10s
2
15s
3
30s 1min
Sampling interval 125ms 250ms 500ms 1s 2s
Selectable Save
Interval Values
10 min to 12 h 10 min to 1
day
10 min to 3
days
10 min to 7
days
10 min to 14
days
Trend/Storage
interval
1
2min 5min 10min 15min 20min
Sampling interval 4s 10s 20s 30s 40s
Selectable Save
Interval Values
10 min to 14
days
10 min to 31
days
10 min to 31
days
10 min to 31
days
1 h to 31 days
Trend/Storage
interval
1
30min 1h 2h 4h 10h
Sampling interval 1h 2h 4h 8h 20h
Selectable Save
Interval Values
1 h to 31 days 1 h to 31 days 2 h to 31 days 4 h to 31 days 8 h to 31 days
1 You can only set a data interval that corresponds to a sampling interval that is slower than
the scan interval.
2 Only available on high-speed input models of the MV.
3 Selectable in fast sampling mode on medium-speed input models of the MV.
4.1 Setting the Recording Conditions of Measured Data