Manual
App-4
M-4660
Appendix 2 Event Action Configuration Examples
Example1:ToStart/StopMemorySamplingbyUsingtheRemoteControl
Function (/R1 Option)
We will configure the MV so that it starts or stops memory sampling when a signal is
applied to remote control input terminal 2. We will use event action number 1.
• DisplayandSettings
Press MENU and then select Menu tab > Timer, Event action > Event action.
<Operation>
If you turn ON the input to remote control input terminal 2, the MV will start memory
sampling (if it isn’t already). If you turn OFF the input, the MV will stop memory sampling.
Example2:ToWriteaMessageWhenanAlarmOccurs
We will configure the MV so that it will write the message “Channel 1 Alarm” to group 1
when an alarm occurs on channel 1. We will use event action number 2.
• DisplayandSettings
Press MENU and then select Menu tab > Timer, Event action > Event action.
<OtherSettings>
• Assign an alarm to channel 1, and configure the MV so that alarm events are sent to
internal switch 3.
• Register “Channel 1 Alarm” to message number 4.
For alarm configuration instructions, see section 3.7.
For message registration instructions, see section 6.4.