Manual
12-6
M-4660
• ErrorsRelatedtotheHistoricalTrend
232 There is no available data. Appears when displaying historical trends.
Specify another file.
233 The specified historical data do not exist. Appears when switching to historical trend from
information display.
Sections 5.7, 5.8, and 5.9
234 The specified channel is not assigned to the display group. Appears when switching to trend, digital, or bar graph
from overview.
See sections 5.4 and 6.1
• ErrorsRelatedtoE-mailandWebServer
260 IP address is not set or ethernet function is not available. The IP address is not specified. Check the IP address.
Section 2.2 in the communication manual
261 SMTP server is not found. Occurs when the SMTP server is specified by name.
• Check the DNS setting.
• Check the SMTP server name.
Sections 2.2 and 2.3 in the communication manual
262 Cannot initiate E-mail transmission. • The host name of the MV is not correct. Check the
host name.
• The port number of the SMTP server is not correct.
Check the port name.
Sections 2.2 and 2.3 in the communication manual
263 Sender’s address rejected by the server. Check the sender address.
Section 2.3 in the communication manual
264 Some recipients’ addresses are invalid. Check the recipient addresses.
Section 2.3 in the communication manual
265 SMTP protocol error. May occur if a network failure (cable problems, duplicate
addresses, network device failure, and so on) occurs in
the middle of the e-mail transmission.
266 Ethernet cable is not connected. Check the cable connection.
Section 2.2 in the communication manual
267 Could not connect to SMTP server. • Check to see that the SMTP server is connected to the
network.
• If the SMTP server name is specified using an IP
address, check to see that the IP address is correct.
Section 2.3 in the communication manual
268 E-mail transmission request failed. Contact your nearest Omega dealer.
269 E-mail transfer error. May occur if a network failure (cable problems, duplicate
addresses, network device failure, and so on) occurs in
the middle of the e-mail transmission.
270 Could not connect to the POP3 server. Check whether or not the destination POP3 server is
running. Check that the cables are connected properly
and that the HUB is turned on. Section 2.3 in the
communication manual
271 POP3 server authentication failed. Check whether the user name and server name are
permitted by the server.
Section 2.3 in the communication manual
275 The current image cannot be output to the Web. The setup display cannot be output to the Web browser.
This message is displayed on the Web browser.
276 Image data currently being created. Unable to perform key
operation.
Try again a little later. This message is displayed on the
Web browser.
277 Could not output screen to Web. Failed to create the image. This message is displayed on
the Web browser.
12.1 List of Messages