Manual

Basic Setting Mode Settings and Default Values
Environment Tab
Operating environment
Setting Selectable Range or Choices Default Value
Operating environment
Tag/Channel Tag/Channel Tag
Language English/Japanese/German/French/Chinese/Korean English
Temperature C/F C
Decimal Point Type Point/Comma Point
View, Message
Setting Selectable Range or Choices Default Value
View
Trend type T-Y T-Y
Partial On/Off Off
Trend rate switching On/Off Off
Message
Write group Common/Separate Common
Power-fail message On/Off Off
Change message On/Off Off
Input, Alarm
Setting Selectable Range or Choices Default Value
Input
Value on over-range Free/Over Free
alarm
No logging On/Off Off
Security, Media save
Setting Selectable Range or Choices Default Value
Security
Key Off/Login/Keylock Off
Communication Off/Login Off
Save
Auto save On/Off On
Media FIFO On/Off Off
Batch
Setting Selectable Range or Choices Default Value
Batch
On/Off On/Off Off
Lot-No. digit Off/4/6/8 6
Auto increment On/Off On
Communication > Service port
Setting Selectable Range or Choices Default Value
Service port
FTP Numeric value (1 to 65535) 21
HTTP Numeric value (1 to 65535) 80
SNTP Numeric value (1 to 65535) 123
Modbus Numeric value (1 to 65535) 502
Default MV1000 Settings
49
IM MV1000-02E