Manual
Quick Settings
Using the quick settings function, you can switch quickly to the measurement channel
(measurement range and alarm) and data save configuration screens by pressing T/DIV.
Procedure
Operate the MV from the Quick settings screen according to the procedure described
below.
• Measurement channels: Set channels 1 through 4 to the following settings:
Mode: Volt, Range: 2V
, Span_L: -2V, Span_U: 2V
•
Display-data save settings: Tre
nd/Storage interval: 1min, Save interval: 1h
• Event data save settings: Sample rate: 1s, Mode: Free (mem. sampling started with
START/STOP key), Data length: 1h
List of default MV1000 settings: page 37
Press MENU.
Press the left arrow key.
Press T/DIV.
Set the measurement range
1. Select the channels to configure.
Press 1 for the first channel and 4 for the last channel.
2. Select the Mode.
Press the Volt soft key.
3. Select the Range.
Press the 2V soft key.
4. Enter the span upper and lower limits.
Press the Input soft key, and select a value with the
up and down arrow keys.
32