Computation and Report Functions (/M1 and /PM1 Options)
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Span Lower, Span Upper
Set the measurement range.
Selectable range of values: –9999999 to 99999999
Selectable decimal places: X.XXXX, XX.XXX, XXX.XX, XXXX.X
Unit
Set the unit of the computed value (Up to 6 characters,
A a # 1
).
Alarm
The available alarm types are high limit alarm, low limit alarm, delay high limit alarm,
and delay low limit alarm.
The range of alarm values is as follows:
Type Value
H, L, T, t Within –9999999 to 99999999 excluding the decimal point
For details on setting alarms, see section 3.7.
* If the Math On/Off or calculation expression is changed, the alarms for that channel are turned
Off.
Alarm delay > Time
Set the alarm delay time using an integer in the range of 1 to 3600 s.
Tag > Comment
Sets the tag comment. On a DX with a release number of 3 or later, you can enter up
to 32 characters. On a DX with a release number of 2 or earlier, you can enter up to
16 characters. The characters that can be entered are:
A a # 1
.
Tag > No. (Release number 3 or later)
This setting only appears when you have enabled the use of tag numbers.
Sets the tag number. (Up to 16 characters:
A a # 1
)
Constant
Number of constant
Select the constant (K01 to K60) to set.
Value
The selectable range is as follows:
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
The number of significant digits of a constant is five. When specifying the
constant using exponential notation, set the mantissa less than or equal to 5
digits and the exponent less than or equal to 2 digits.
TLOG
Timer type
You can set the timer type to Timer or Match T (match time timer).
Timer No.
Select the timer number to use.
For details on setting the timer, see section 7.1.
For match time timer setting instructions, see section 7.1.
Sum scale
Set the sum scale to /s to /h to match the unit of the measured value.
Example: If the unit of the measured value is “m
3
/min,” select /min.
Off: Sums as-is the measured data per scan interval.
Reset
To reset the TLOG computed value at each interval, select On.
9.1 Setting the Expression, Measurement Range, Alarm, Tag, and Data Storage on Computation Channels