Owner's manual

62
IM 04L42B01-02E
Connecting to the RS-422/485 (/C3)
FG (Frame Ground) Frame ground of the DX.
SG (Signal Ground) Signal ground.
SDB (Send Data B) Send data B (+).
SDA (Send Data A) Send data A (–).
RDB (Received Data B) Receive data B (+).
RDA (Received Data A) Receive data A (–).
FG SG SDB SDA RDB RDA
Electric
potential
of the shield
Two-wire system
Shield
FG SG SDB SDA RDB RDA
Electric
potential
of the shield
Four-wire system
Shield
Recommended length of stripped wire: 9 mm. Recommended tightening torque: 0.4-0.5 N•m.
Cable
There are two types of cables available, the four-wire cable and the two-wire cable (used only for the Modbus protocol).
The cable should meet the following specifications.
Type: Shielded twisted pair cable: 3 pairs 24 AWG or more (four-wire),
2 pair 24 AWG or more (two-wire)
Characteristic impedance: 100
Capacitance: 50 pF/m
Total cable length: Up to 1.2 km
Connecting to the VGA Connector (/D5)
CAUTION
Connect the cable after turning OFF the DX and the monitor.
Do not short the VIDEO OUT terminal or apply external voltage to it. If you do,
the DX may malfunction.
VGA Output Terminal
!
VIDEO OUT(VGA)
The DX display can be output to a monitor through the RGB output. Connectable
monitors are VGA monitors or multi-sync monitors capable of displaying VGA.
Specifications of the VGA Output Terminal
Pin No. Signal Name Specifications
1 Red 0.7 Vp-p
2 Green 0.7 Vp-p
3 Blue 0.7 Vp-p
4
5
6 GND
7 GND
8 GND
9
10 GND
11
12
13 Horizontal sync
signal
Approx. 31.5 kHz, TTL
negative logic
14 Vertical sync signal Approx. 60 Hz, TTL
negative logic
15
Installation and Wiring
6
11
1
5
15
10
D-Sub 15-pin receptacle