User Manual

IM 01E20C02-01E
6-27
6. PARAMETER DESCRIPTION
Output Example 2
The high alarm (H) is set to 80% or more of the flow
rate span; the low-low alarm (LL), to 20% or less.
Settings are:
G10: Low Alarm = -110%
G11: High Alarm = 80%
G12: Low Low Alarm = 20%
G13: High High Alarm = 110%
0 Time (t)
H
DO
DIO
LL
Instantaneous flow rate
20
50
80
100
[%]
Select “H/L Alarm(O)” for F20: DO Function
Select “HH/LL Alarm(O)” for F21: DIO Function
Select “Closed (On) Act” for F22: DO Active Mode
Select “Closed/Short Act” for F23: DIO Active Mode
F0609.EPS
L Alarm set to
-110% indicates
that the alarm is
disabled.
HH Alarm set to 110% indicates
that the alarm is disabled.
High Alarm
Low Low Alarm
Output Example 3
The high alarm (H) is set to 80% or more of the flow
rate span; the high-high alarm (HH), to 90% or more.
Settings are:
G10: Low Alarm = -110%
G11: High Alarm = 80%
G12: Low Low Alarm = -110%
G13: High High Alarm = 90%
0 Time (t)
H
HH
DO
DIO
Instantaneous flow rate
50
80
90
100
[%]
F0610.EPS
High High Alarm
High Alarm
Select “H/L Alarm(O)” for F20: DO Function
Select “HH/LL Alarm(O)” for F21: DIO Function
Select “Closed (On) Act” for F22: DO Active Mode
Select “Closed/Short Act” for F23: DIO Active Mode
L Alarm set to
-110% indicates
that the alarm is
disabled.
LL Alarm set to -110% indicates
that the alarm is disabled.
NOTE
Although the same items can be selected using
the DO terminal (selected for F20) and the DIO
terminal (selected for F21), output is identical
for both.
Setting values of -110% and 110% are used to
disable corresponding functions; and accord-
ingly, status output can be customized for
specific purposes.