User Manual
IM 01E20F12-01E
9-7
9. PARAMETER LISTS
T0903-3.EPS
Index Parameter Valid Range Initial Value Description
Write
Mode
77 MASS_FLOW_DENSITY Setting of the density for mass-flow rate. This parameter is necessary
in situations where t, kg, g, klb or lb has been selected as the mass
unit in VOLUME_FLOW_UNITS. If a mass unit is selected in
VOLUME_FLOW_UNITS and a value of 0 is set for this parameter,
the setting alarm “57: Density SetErr.” will be displayed. In such a
case, ensure that the density is set correctly.
O/S 0 to 32000 0
78 LIMSW_1_
VALUE_D
(DS-34)
This parameter is used in order to indicate the current input value
selected LIMSW_1_TARGET. If the target alarm occurred the value
is set 1. Refore to LIMSW_1_TARGET.
– 0: not set,
set exclude 0
0
Value
Value
79 LIMSW_1_TARGET This parameter selects the input channel used to LIMSW_1_
VALUE_D.
O/S 0: Volume Flow
Value
0
80 LIMSW_1_SETPOINT Sets the threshold of limitswitch 1. If the value of
LIMSW_1_ACT_DIRECTION is HIGH LIMIT, limit switch 1 turns ON
when LIMSW_1_TARGET has gone beyond LIMSW_1_SETPOINT.
If the value of LIMSW_1_ACT_DIRECTION is LO LIMIT, limit switch
1 turns ON when LIMSW_1_TARGET has gone below
LIMSW_1_SETPOINT. The unit set in LIMSW_1_UNIT applies.
O/S -10
81 LIMSW_1_ACT_
DIRECTION
Selects the direction of the limit switch 1’s actions:
1: LO LIMIT (Low-limit switch)
2: HIGH LIMIT (high-limit switch)
O/S 0: Low Limit0: Low Limit
1: High Limit
82 LIMSW_1_HYSTERESIS Sets the hysteresis of limit switch 1 to be applied for resetting the
LIMSW_1_VALUE_D to OFF after LIMSW_1_TARGET went beyond
LIMSW_1_SETPOINT and LIMSW_1_VALUE_D turned
ON (when used as a high-limit switch), or after LIMSW_1_TARGET
went below LIMSW_1_SETPOINT and LIMSW_1_VALUE_D turned
ON (when used as a low-limit switch).
O/S 0.00 or larger.
83 LIMSW_1_UNIT Indicate the unit set in LIMSW_1_TARGET.
– 1349: m
3
/h
84 LIMSW_2_
VALUE_D
(DS-34)
This parameter is used in order to indicate the current input value
selected LIMSW_2_TARGET. If the target alarm occurred the value
is set 1. Refore to LIMSW_2_TARGET.
–00: not set,
set exclude 0
85 LIMSW_2_TARGET Select the input channel to LIMSW_2_VALUE_D.O/S 00: Volume Flow
Value
86 LIMSW_2_SETPOINT Sets the threshold of limitswitch 2. If the value of LIMSW_2_ACT_
DIRECTION is HIGH LIMIT, limit switch 2 turns ON when LIMSW_
2_TARGET has gone beyond LIMSW_2_SETPOINT. If the value of
LIMSW_2_ACT_DIRECTION is LO LIMIT, limit switch 2 turns ON
when LIMSW_2_TARGET has gone below LIMSW_2_SETPOINT.
The unit set in LIMSW_2_UNIT applies.
O/S 10
87 LIMSW_2_ACT_
DIRECTION
Selects the direction of the limit switch 2’s actions
0: LO LIMIT (Low-limit switch)
1: HIGH LIMIT (high-limit switch)
O/S 1: High Limit0: Low Limit
1: High Limit
88 LIMSW_2_HYSTERESIS Sets the hysteresis of limit switch 2 to be applied for resetting the
LIMSW_2_VALUE_D to OFF after LIMSW_2_TARGET went beyond
LIMSW_2_SETPOINT and LIMSW_2_VALUE_D turned ON
(when used as a high-limit switch), or after LIMSW_2_TARGET
went below LIMSW_2_SETPOINT and LIMS W_2_VALUE_D turned
ON (when used as a low-limit switch).
O/S 00 or larger.
89 LIMSW_2_UNIT Indicate the unit set in LIMSW_2_TARGET.– 1349: m
3
/h
90 SWITCH_1_
VALUE_D
(DS-34)
Indicate the value of switch 1, which switches ON and OFF
depending on the digital value of the target input parameter selected
in SWITCH_1_TARGET.
–00: not set,
set exclude 0
91 SWITCH_1_TARGET This parameter selects the input channel used to LIMSW_1_
VALUE_D.
O/S 0: Adheshion
Alarm
0: Adheshion Alarm
1: Adhesion Warning
–Good
Status
–Good
Status
Value
–Good
Status