7. DIAGNOSTIC INFORMATION
Table 7.2 List of DIAGNOSIS_EXTENSION
T0702.EPS
0 10:uP Fault Microprocessor error
1 11:EEPROM Fault
2 12:A/D(H) Fault Contact nearest office or service center A/D converter error
3 13:A/D(L) Fault
4 14:A/D(Z) Fault
5 15:Coil Open Cut the power and check coil & EX cables Flowtube coil open-circuit
6 16:EEPROM Dflt EEPROM default values
7 100:Comm uP Fault Contact nearest office or service center FB Microprocessor error
0 101:Comm EEPROM Fault FB EEPROM error
1 102:Total1 Not Saved Total1 had not been stored when the device was
turned off.
2 103:Total2 Not Saved Total2 had not been stored when the device was
turned off.
3 104:Total3 Not Saved Total3 had not been stored when the device was
turned off.
4 105:Comm Error1 Contact nearest office or service center Communication error
5 106:Comm Error2 Contact nearest office or service center Communication error
6 (Reserved)
7 (Reserved)
0 (Not Used)
1 (Not Used)
2 (Not Used)
3 (Not Used)
4 30 Sig Overflow Check signal cable and grounding Input signal error
5 31:Empty Pipe Fill flow tube with fluid Flowtube is not filled with fluid
6 33:Adhesion Alm Clean electrodes Electrode adhesion alarm
7 (Reserved)
0 50:Span > 10m/s
Check parameter 27 of AI and 34 of TB
Span flow velocity setting is 11 m/s or more
1 51:Span < 0.1m/s Span flow velocity setting is 0.05 m/s or less
2 57:Dens Set Err
Check parameter 34, 76 and 77 of TB
Mass units have been selected for Base Flow
Unit but density is set to 0.
3 71:Meas Mod Set
Check parameter 70 of TB
Measure_Mode is set to Enhanced DF without
selecting an optional specification code /HF1 or /HF2.
4 72:Size Set Err
Check parameter 31 and 32 of TB
A value of 3000.1 mm or more is set for
Nominal_Size.
5 73:Adh Set Err
Check parameter 115 to 118 of TB
The condition in Adhesion detection level,
Level:1<Level:2<Level:3<Level:4 is not satisfied.
6 (Not Used)
7 (Not Used)
0 (Not Used)
1 (Not Used)
2 (Not Used)
3 (Not Used)
4 80:Adhesion Wng Slight adhesion to electrodes.
5 (Reserved)
6 82:Auto Zero Wng Results of automatic zero adjustment are higher
than the rated values.
7 85:Flow Vel Over Flow velocity overflow.
0 (Reserved)
1 (Not Used)
2 (Not Used)
3 (Not Used)
4 (Not Used)
5 (Not Used)
6 (Not Used)
7 140:Sim.Jmpr On Simulation Jumper is set to ON (Hardware).
System
Alarms
1
2
3
4
5
6
(Not Used)
(Not Used)
(Not Used)
Warning
Warning
Process
Alarms
Setting
Alarms
Category
DIAGNOSIS_
EXTENSION
Display Message Countermeasure Message
Alarm Description
Octet Bit