4-7
4. CONFIGURATION
Table 4.4 TOTAL
Quality (bit 6-7) Sub-status (bit 2-5) Limit (bit 0-1) Alarm Value
T0404.EPS
2:Good (NC) 0:Ok 0:ok
1:Update Event
2:Active Advisory Alarm 1:low limited --:Total1 Lo Alm 0x89
--:Total2 Lo Alm
--:Total3 Lo Alm
2:high limited --:Total1 Hi Alm 0x8A
--:Total2 Hi Alm
--:Total3 Hi Alm
3:Active Critical Alarm: 1:low limited --:Total1 Lo Lo Alm 0x8D
--:Total2 Lo Lo Alm
--:Total3 Lo Lo Alm
2:high limited --:Total1 Hi Hi Alm 0x8E
--:Total2 Hi Hi Alm
--:Total3 Hi Hi Alm
4:Unack. Update Event
5:Unack. Advisory Alarm
6:Unack. Critical Alarm
9:Maintenance required 0:ok 80:Adhesion Wng 0xA4
81:(Reserved)
1:Uncertain 0:Non specific 0:ok 10:uP Fault 0x40
11:EEPROM Fault
12:A/D(H) Fault
13:A/D(L) Fault
14:A/D(Z) Fault
16:EEPROM Dflt
100:Comm uP Fault
101:Comm EEPROM Fault
105:Comm Error1
106:Comm Error2
15:Coil Open
30:Sig Overflow
31:Empty Pipe
33:Adhesion Alm
34:(Reserved)
41:TB O/S Mode
50:Span > 10m/s
51:Span < 0.1m/s
57:Dens Set Err
71:Meas Mod Set
72:Size Set Err
73:Adh Set Err
82:Auto Zero Wng
1:low limited or 85:Flow Vel Over 0x41
2:high limited 0x42
0:Bad 1:Configuration Error 0:ok 120:Total1 Unit Set Err 0x04
121:Total2 Unit Set Err
122:Total3 Unit Set Err
7:Out of Service 3:constant 40:PB O/S Mode 0x1F
43:Total1 FB O/S Mode
44:Total2 FB O/S Mode
45:Total3 FB O/S Mode