5.3 Insulation
The heat exchanger and top ange section of the YARS MUST always be insulated by at least two inches (obtained from
a local supplier). This will reduce the inuences of ambient air temperature uctuations and cause the heat exchanger to
operate more eciently. Insulation should be installed at site after installation and leak testing and should be congured for
possible future maintenance and access.
It is generally required to have the lter section insulated only when ambient temperatures fall below 32
o
F (0
o
C). Insulation
should conform to appropriate site standards. This insulation may not be required during hot summer months however
only on-site experience can determine the exact insulation requirements for the lter section.