User guide
25
IM 11M06D03-01E-A
Zero Pressure Switch, M1133AR 
 ThepressureswitchisusedtoalerttheOperatorviaacontactinputthatthezerogascylinderneedstobere-
placedbeforethecylinderisempty.Therecommendedconnectionistothehighpressuresideofthezerogas
regulator. Refer to Figure 4.1.
Zero Pressure Switch, M1133AR
AdjustableSetPointRanges:   30to575psionfall;50to600psionrise.
DeadBand:     8to60psi
OverRange:     2,500psi
ProofPressure:    3,000psi
StorageTemperature:    -40ºFto180ºF(-40ºCto82ºC)
AmbientTemperature:    0ºFto160ºF(-18ºCto71ºC)
SetPointRepeatability:   1%ofspan
SwitchOutput:     OneSPDT(eitherNCorNO)
ElectricalRating:    Ratedto5ampsresistiveand15ampsinductive(75%PF), 
      at125and120VAC,¼”HP
Enclosure:     Aluminumwithirradiatefinishratedfor100hoursofsalt 
  spray
ElectricalConnection:    OneSPDToutput;½”NPTF,5ftcable
PressureConnection:    1/8”NPTM
Mounting:     NPTMpressureconnection
Weight:     12oz(340g)
Approvals:     UL508listed,E42272;CSAC22.2,No14-1987:LR9690