User guide
25
IM 11M06D03-01E-A
Zero Pressure Switch, M1133AR
ThepressureswitchisusedtoalerttheOperatorviaacontactinputthatthezerogascylinderneedstobere-
placedbeforethecylinderisempty.Therecommendedconnectionistothehighpressuresideofthezerogas
regulator. Refer to Figure 4.1.
Zero Pressure Switch, M1133AR
AdjustableSetPointRanges: 30to575psionfall;50to600psionrise.
DeadBand: 8to60psi
OverRange: 2,500psi
ProofPressure: 3,000psi
StorageTemperature: -40ºFto180ºF(-40ºCto82ºC)
AmbientTemperature: 0ºFto160ºF(-18ºCto71ºC)
SetPointRepeatability: 1%ofspan
SwitchOutput: OneSPDT(eitherNCorNO)
ElectricalRating: Ratedto5ampsresistiveand15ampsinductive(75%PF),
at125and120VAC,¼”HP
Enclosure: Aluminumwithirradiatefinishratedfor100hoursofsalt
spray
ElectricalConnection: OneSPDToutput;½”NPTF,5ftcable
PressureConnection: 1/8”NPTM
Mounting: NPTMpressureconnection
Weight: 12oz(340g)
Approvals: UL508listed,E42272;CSAC22.2,No14-1987:LR9690