8.2.3 SPAN-GAS CONCENTRATION
Settheoxygenconcentrationforspancalibration.Ifinstrumentairisusedasthespangas,enter21%O2.
1) Select Span gas conc. from the Calibration setup display.
2) Enterthedesiredspan-gasoxygenconcentrationfromthenumeric-dataentrydisplay.(Thespan-gasset
rangesfrom4.5to100%O2.)
3) Enter02100foranoxygenconcentrationof21vol%O2.Instrumentairisheredenedasdryairwitha
dew-pointtemperatureofnohigherthan-208˚C.
Ifthedew-pointtemperatureishigherthan-208˚C,useahand-heldoxygenanalyzertomeasuretheactual
oxygen concentration.
(1) When instrument air is used for the span calibration, remove the moisture from the instrument air at a
dew-point temperature of -208C and also remove any oily mist and dust from that air.
(2) If dehumidifying is not enough, or if foul air is used, the measurement accuracy will be adversely affected.
8.3 MANUAL CALIBRATION
For manual calibration, consult Section 7.12, Calibration, In the ZR22G / ZR402G Instruction Manual (IM 11M12A01-02E)
8.4 SEMI-AUTOMATIC CALIBRATION
Tostartcalibration,followthesesteps:
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
(2) ThenselectCalibrationfromtheExecution/Setupdisplay.TheCalibrationdisplayshowninFigure8.4
appears.
(3) (2)SelectSemi-autocalibrationtodisplaytheSemi-automaticcalibrationdisplayshowninFigure8.4
(4) (3)SelectStartcalibration.ThedisplayshowninFigure8.4appears,andthenstartcalibration.
Figure 8.4
CAUTION