7.3 CHECKING FOR LEAKS
Topreventleakage,allthreadedconnectionsshouldhaveTeontape(orsuitablealternative)andall
compressionttingsshouldbeinstalledpermanufacturer’srecommendations.
1. LocatetheBLOCKSOLENOID.TurntheoverridescrewontheBLOCKSOLENOIDandSPANSOLENOIDtothe
  manual position.
2. Useleakdetectionsprayonallcompressionttingsandbendsofthecalandreferencelinetubing.
3.  Inspect the full length of the tubing to determine if there is a leak.
4.  Repair leaks as necessary, and within applicable guidelines.
5.  After repairing the leaking areas, repeat the inspection instructions above for the affected areas.
6. Uponcompletionofrepairingtheleaks,returnallsolenoidstotheAUTOposition.