User guide

13
IM 11M06D03-01E-A
VI. INSTALLATION, AUTOMATIC CALIBRATION, AC1
6.1LOCATION
Thefollowingguidelinesshouldbeusedwhenselectingalocationforthecalibrationunit:
1) Easily accessible for maintenance and inspections.
2) AsclosetotheZR22probeaspractical.Thiswillminimizetheamountoftubingrequiredforplumbing,and
decreasecalibrationtimes.Ambienttemperatureshouldnotexceed55˚C(131˚F).
3) Humidityismoderateandnocorrosivegasesarepresent
Use an air purge for the AC1 enclosure if the corrosive gases or high dust content is present.
4) Minimal vibration
5) Clean, dry instrument air is available
6.2MOUNTINGTHEAC1
TheAC1isavailableinaNEMAandNEMA4X,lockableenclosure,andtheunitisdesignedforwallmounting.Itcan
be secured using four (4) standoffs and four (4) bolts. Mount the AC1 unit so that the terminal strip is easily visible for
wiringpurposes.Sufcientroomshouldbemadeavailableforconnectingthecalibrationandreferencegastubing.
Theunitmustbemountedaslevelaspossibletoensuretheaccuracyoftheowrates
6.3PIPING
AlistofrecommendedpartforpipingtotheMC1andAC1unitsislistedpreviouslyinTable4.1
6.3.1PIPINGTOTHEZR22GOXYGENPROBE
6.3.2CONNECTIONTOTHECALIBRATIONGASINLET
ConnectthetubingfromtheZero(cal)gasoutput,showninFigure5.1,totheCALINsideoftheZR22probe.Ifthe
processisapositivepressureapplication,ensurethatacheckvalve(YokogawaPartNumberM1234VV-A)is
installedbetweentheCALINportandthe1/4”tubing.Thecheckvalvemayalreadybeinstalledonthedetector
prior to shipping.
6.3.3CONNECTIONTOTHEREFERENCEGASINLET
ConnectthetubingfromtheReferencegasoutput,showninFigure5.1,totheREFINsideoftheZR22probe.
Clean,dryinstrumentairisrecommendedasareference(span)gas;however,ambientairmaybeusedasa
substitute.
6.4.WIRINGTOTHEAC1,AUTOMATICCALIBRATIONUNIT
6.4.1WIRINGTHEAC1TOTHEZR22G
ThesolenoidsoftheAC1arepoweredby24VDCfromtheZR402GOxygenConverter.TheAC1itselfcanbe
poweredbytheZR402GiftheZR402Gisusing120VAC,60Hz.SupplyvoltagefortheAC1canbeobtainedfrom
terminals“L,N&G”ontheZR402G.PleaserefertothewiringdiagramlocatedinAppendixC.
NOTE
NOTE