4.3.3 CONNECTION TO THE REFERENCE GAS INLET
ConnectthetubingfromtheReferencegasoutput,showninFigure4.3,totheREFINsideoftheZR22probe.Clean,
dryinstrumentairisrecommendedasareference(span)gas;however,ambientairmaybeusedasasubstitute.
Figure 4.1
AlistofrecommendedpartforpipingtotheMC1orAC1unitsislistedbelowinTable3.1
Calibration Unit  Piping 
Location 
Parts  Notes 
Check valve 
Recommended by Yokogawa for positive 
pressure applications: M1234VV-A. 
Available through Yokogawa 
Nipple  Rc 1/4 or 1/4 inch MNPT 
General 
parts 
Zero Gas Cylinder  User’s Scope 
Gas Pressure 
Regulator 
Recommended by Yokogawa: M1132KD 
Regulator, and M1132ME Gauge. 
Available through Yokogawa 
Calibration Gas 
Inlet 
Zero Gas Pressure 
Switch 
Recommended by Yokogawa: M1133AR 
Available through Yokogawa 
MC1 Manual Calibration 
Unit 
AC1 Single Channel 
Automatic Calibration 
Unit 
Reference gas 
inlet 
Joint for tube 
connection 
Rc 1/4 or 1/4 inch MNPT 
General 
parts 
INSTRUMENT AIR
INPUT
ZERO (CAL)
GAS INPUT
REFERENCE AIR
OUTPUT
CALIBRATION 
GAS OUTPUT
Figure 4.1