IV. WIRING
4.1 Wiring the AC4/AC8 to the AV550G Averaging unit
Please refer to appendix A for the system conguration 
drawing.
Solenoid, Relay and Pressure Switch wiring connections
The AC4/AC8 uses 120VAC, 60 Hz for all solenoids. Al-
though two internal 1 amp fuses are provided for the ±120 
VAC terminal, an external 5 amp circuit breaker is recom-
mended. Included with each fuse is an LED that is activated 
when the fuse is blown. Recommended wiring is 14AWG, 
run in conduit. If conduit is not used, wiring should have a 
suitable jacket to meet environmental and regulatory codes. 
The AC4 unit requires a maximum of 8 conductors, whereas 
the AC8 requires 12 conductors.
NOTE
A 120 VAC relay (included with NEMA 4 enclosure), shown at ter-
minal K1, is required at the AV550G if a pressure switch is used. 
This prevents noise caused by a low voltage pressure switch 
signal from being run in the same cable as the 120 VAC solenoid 
signal.
The relay is prewired as NORMALLY CLOSED (NC) to 
contact input #1 at the control card (terminal 7 and 9) of the 
AV550G.
 Figure 4.1 AV550G Input Contact Screen
WARNING
When the relay is deactivated, the unit will not calibrate. The 
relay remains closed unless a 120 VAC source is connected to 
energize it. Terminal 7 of the relay provides 120 VAC. Neutral 
voltage is connected to terminal 2 of the relay from the AC4/AC8 
calibration unit.
The relay switch is designed to OPEN when energized as 
depicted in the ow diagram shown in Figure 4.2. A pressure 
switch input completes the 120 VAC signal used to energize 
the relay, which is wired at the calibration unit as Normally 
Open (NO), where Open = Low or Zero tank pressure. The 
AC4/AC8 unit is prewired with the jumper in place, so that 
the relay will be energized and OPEN. However, if the pres-
sure switch input is used, then remove the jumper between 
the terminals labeled PS at the calibration unit’s terminal 
strip, and attach the pressure switch. The pressure switch is 
wired as normally open.
Pressure
Switch
Relay
Calibration
Status
Cal Gas
Pressure
OK ?
Relay Deen ergized
and
CLOS ED
Pressure Switch
OPEN
Relay Ener gi z ed a nd
OPEN
Pressure Switch
CLOS ED
CALIBRATE
YES
NO
STAR T
DO NOT
CALIBRATE
Pressure Switch U sed a s I np ut  to R elay
Calibration
Using
Pressure Switch AND Relay
Figure 4.2: Calibration Using Pressure Switch and Relay Flow 
Chart
The recommended pressure switch conguration is to the 
high pressure side of the zero gas regulator. This is preferred 
because it gives the earliest warning of a decrease in zero 
gas pressure. The switch is threaded into the high pres-
sure purge port of the dual stage regulator on the zero gas 
cylinder.
4.2 Pressure Switch, M1133AR, Set Point Procedure
For the AC4/AC8 Autocal units, a gas pressure set point is 
used to indicate that the calibration gas (zero gas) is low. 
Pressure Switch M1133AR is a cylindrical style pressure 
switch which uses a diaphragm or piston sensor to detect a 
pressure change. The recommended set point is 200 psi on 
the fall. At the predetermined set point, a SPST switch actu-
ates, converting a pressure signal into an electrical signal.
Figure 4.3: Pressure Switch and Dual Stage Regulator
The following procedure is required to adjust the set point:
1. Thread the pressure switch into the high pressure gauge 
port on the regulator.
2. With power disconnected, slide the cover on the switch 
toward the electrical termination while twisting it to over-
come friction.
3. Connect the power to the terminal or leads.
4. Insert a at head screwdriver (1/8”) into the adjustment 
slot and turn counterclockwise to increase the setting or 
clockwise to decrease the setting.
M1133AR
Pressure Switch
M1132ZX Dual 
Stage Regulator