Manual
5
IM 11M6D2-02E-A
3.4 PIPING
PCDE?Q
A?J
E?Q
/-2ĥ
DCK?JCLNR
/-2ĥDCK?JCLNR
WMIME?U?
O2 CALIBRATION SYSTEM
LCK?2V
INSTRUMENT AIR 
ZERO GAS 
35 psig MAX 
PROBE 1
PROBE 2
PROBE 3
PROBE 4
PROBE 5
PROBE 6
PROBE 7
PROBE 8
Figure 3.2
Piping Gases to the Auto Cal Unit
Piping (or tubing) is required from the instrument air line and 
zero gas cylinder to the calibration unit. Standard tubing is _” 
O.D. (Outside Diameter) copper or stainless steel. Stainless 
steel is preferred, as it is less likely to develop leaks. The 
separate tubing for both the instrument air and zero gas are 
connected to the two separate 1/4” FNPT brass or stainless 
ttings on the LEFT side of the calibration unit. 
Piping Gases to ZR22 Probes
The tubing to the ZR22 probes is run from the RIGHT side of 
the calibration unit to each of the ZR22 probes as shown in 
Figure 2. Quarter inch (1/4”) tubing should be used between 
the auto cal unit and the probes. It is important to note 1) 
the reference air tubing is connected to the REF IN port 
at the probe; 2) cal gas tubing is connected to the CAL 
IN port on the probe. The REF IN and CAL GAS ports are 
located on the bottom of the ZR22 probe junction box. The 
check gas and reference air ports require a 1/4” FNPT tting.  
NOTE
Flexible stainless steel hose is recommended for the connection 
between the ZR22 and the tubing.
This will allow for probe movement due to thermal expansion of 
the duct.
IMPORTANT
To prevent leakage, all compression ttings should be installed 
per manufacturer’s recommendations. In addition to this, a 
check valve is usually installed on the cal gas inlet of the ZR22 
probe to protect the cal tubing from moisture contamination.










