<6. OPERATION OF SC (Conductivity)>
6-7
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
n Contact status
This screen is displayed only if the PH201G distributor is used and “PH201G” is selected in
communication setup.
n c.c. (factory)
This parameter displays the cell constant (factory setting).
The cell constant (factory setting) is determined by factory calibration made during sensor
manufacturing.ThisvaluecanbesetinCommissioning→Measurementsetup→Sensorsetup.
The cell constant is indicated on the sensor or cable label.
n c.c. (adjusted)
This parameter displays the cell constant (adjusted).
The cell constant (adjusted) is a cell constant that has been calibrated and is set by calibration
action.
Whenthesystem’scellconstantisadjustedon-linebycalibrationusingaprocesssolutionor
buffer solution, a new cell constant is logged here. The difference between this value and the
default value set at factory shipment will not be large. If there is a large difference between them,
check whether the sensor is broken or contaminated.
n Temp. comp. 1
This parameter shows the temperature compensation method for 1st measurement. The setting
ismadeinCommissioning→Measurementsetup→Temp.comp.
n Temp. comp. 2
This parameter shows the temperature compensation method for 2nd measurement. The setting
ismadeinCommissioning→Measurementsetup→Temp.comp.
2nd measurement does not imply that two measurements can be made individually, but it means
thattwotypesofcompensationmethodscanbecongured,whichenablestwophasestobe
monitoredaccuratelyinthesameprocess.Anexampleismonitoringtheswitchingbetween
process and washing solutions.
n Polarization
Thisparameterindicatesthedegreeofpolarizationofthesensor.Itisanindexofthedegree
of sensor contamination (adherents). The greater the value, the more the polarization has
progressed. If the value is high, maintenance should be performed on the sensor as soon as
possible.
n Sensor Ohm
This parameter represents the non-compensated resistance of the sensor.
n USP
USPstandsforUnitedStatesPharmacopeiaandindicatesthemarginofsafetyforthewaterfor
injection (WFI) conductivity at each temperature as a percentage.