User Manual

<4. COMMISSIONING OF pH/ORP>
4-15
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
n
HART
Select this menu when HART communication is made.
IntheHARTsetupscreen,specifythenetworkaddressandsetupparametersforSV,TV,and
FV.
(PVislinkedwiththe“processparameter”settingin“Outputsettings”andcannotbechanged
here.)
l Network address
For1-to-1communication,leavethedefaultvalue[0]unchanged.Formulti-dropwheremultiple
HART devices are connected on a bus, set addresses in 1 to 15. In this case, the mA output will
bexedto4mA.
l PV
PVisaparameterselectedforanalogoutput;itcannotbechangedhere.
l SV, TV, FV
TheSV,TV,andFVparametersareitemsthattheusermustsetup.Selectableitemsdiffer
depending on the sensor type in “Sensor setup” and settings in “Measurement setup.”
If blank is selected for a parameter, items below that parameter must all be set to blank. If an item
is blank, those below it cannot be set to a status other than blank.
For more information on HART communication, see the Technical Information (TI 12A01A02-
60E)intheattachedCD-ROM.
n PH201G
Select this menu if the PH201G distributor is connected to the instrument.
In the PH201G setup screen, make settings for “Hold contact,” “Fail contact,” and “Wash
contact.”
l Hold contact
SelectDisabledorEnabled.
When this item is enabled, the output will be held according to the setting of “Hold type” on the
Hold setup screen.
l Fail contact
Selectastatusfromamong“Fail+Warn,”“Failonly,”and“Disabled.”
Thissettingdependsontheerrorconguration.See“4.4Errorconguration.”
“Fail” corresponds to “Fault.”
PH