User Manual
<4. COMMISSIONING OF pH/ORP>
4-12
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
Table4.2 Errorconguration
Display item Description Default
pH too high ThepHvalueexceeds16.00. Warn.
pH too low The pH value is lower than –2.00. Warn.
Temperature too high Measured process temperature is higher than the
maximumlimit.
Warn.
Temperature too low Measured process temperature is lower than the
minimum limit.
Warn.
ORP too high TheORPvalueishigherthan1500mV. Off
ORP too low TheORPvalueislowerthan–1500mV. Off
rH too high The rH value is higher than 100. Off
rH too low The rH value is lower than 0. Off
Matrixcong.error Thetemperaturecompensationmatrixisnotset
properly (see Section 4.2.3).
Fault
Calib.timeexceeded Calibrationtimeexceedsthecalibrationinterval
(see Section 4.2.4).
Off
Wash half-time error The response to wash is abnormal. Off
Impedance 1 too high The sensor must be checked. Off
Impedance 1 too low
Impedance 2 too high
Impedance 2 too low
CAUTION
Ifcancelinganerrorcongurationcouldberisky,donotcancelitasadangeroussituationmay
result.
NOTE
The “Impedance 1 too high/too low” and “Impedance 2 too high/too low” error items are displayed
intheErrorcongurationscreen.Ingeneral,forapHsensor,“Impedance1”means“glass
electrode,” and for an ORP sensor, it means “metal electrode.” “Impedance 2” represents the
reference electrode in all cases. The “Impedance 1 too high/too low” and “Impedance 2 too high/
toolow”errorsareappliedtosensorswithaliquidearth.Sensorswithaliquidearthinclude
PH8EFP,PH8ERP,PH8EHP,OR8ERG,OR8EFG,etc.Forsensorsnotequippedwithaliquid
earth,theseerrorsshouldallbesettoOff;thesesensorsincludeHA405,HA406,DPA406,
DPA405,DPAS405,HF405,HA485,DPA485,etc.Moreover,formeasuringasolutionwithlow
conductivity such as when using a pH meter for pure water, set the “Impedance 2 too high” error
to Off. Otherwise, no impedance can be detected.
NOTE
WhenORPisspeciedasasensortypein“Sensorsetup”,andwhenanORPsensor,suchas
OR8ERGandOR8EFG,doesn’thaveanytemperatureelement,setbothof“Temperaturetoo
high/low”in“Error”toOffsothatnorelatederroriscaused.Duetotheopeninputoftemperature
signal, a false error may happen as if temperature goes over a high limit or a low limit.