User Manual
<3. OPERATION OF pH/ORP>
3-10
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
History information on events are automatically stored on the preset conditions.
Inadditiontothisstoring,followingthreemessagescanbemanuallystoredinthelogbook;
"Sensor washed by hand", "Module replaced", "Sensor replaced"
To store these messages, press
and select one of three messages from the Item on the
Memorandum screen. Its event date/time will be the time when a message is selected and
entered.
WhenapasswordforCommissioningissetonthepasswords’setup,pressing
requires
entering the password. (Refer to the section 4.6.3.)
Read Logbook:
logbook1-1
Memorandum:
Item
Sensor washed by hand
Enter?
No
logbook1-1 1/1
Memorandum:
Item
Sensor replaced
Enter?
No
2010/02/15 17:04
Sensor replaced
2010/02/15 15:15
Power on
Finish
2010/02/15 15:15
Power on
Finish
logbook1-1 1/1
** When storing message manually.
**
(Fig.3.6)
Sensor washed by hand
Module replaced
Sensor replaced
logbook1-1
logbook1-2
No
Yes
Figure 3.9 Detail display (continued)
3.5 Trend graphics
Pressing on the Zoom display changes the display to a graphical mode in which the
average measured value is shown on a time scale. The “Live” value is also digitally displayed in
atextbox.Thetimescale(X-axis)andtheprimaryvaluescale(Y-axis)aresetinthe“DISPLAY
SETUP”menu(Section4.7.2).
The screen displays the trend of up to 41 averages of the measurement for each time interval.
The FLXA21 samples the measurements every second. The trending graphic also shows the
maximumandminimummeasuredvaluesinthatinterval.
Forexample,ifthetimescaleissetto4hours,thenthetrendisshownfor4hourspriortothe
actualmeasurement.Eachpointonthetrendlinerepresentstheaverageover4×60×60/41=
351 measurements (seconds).