User Manual

<3. OPERATION OF pH/ORP>
3-9
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
As shown in Figure 3.8, the date is predicted based on the intersection point of the upper or lower
limitsandtheextrapolatedlineofthevaluesobtainedbytheleastsquaresmethod.
Day
Value
Lower limit
Upper limit
Current
date
Projected
maintenance date
F0310.ai
Figure 3.8
The status shows the certainty of the projected maintenance date in terms of the correlation
coefcientR.Tables3.1and3.2showrespectivedisplaypatterns.
Table 3.1 Display pattern of the projected maintenance date
Projected date - - 0-1 month 1-3 months 3-6 months 6-12 months Over 1 year
--: cannotbepredictableduetoinsufcientdata
Table 3.2 Display pattern of the status
Status
(- - - - -)
(R < 0.50)
(Poor)
(0.50≤R<0.70)
(Reasonable)
(0.70≤R<0.85)
(Excellent)
(0.85≤R<1.00)
n Projected replacement
The projected replacement function predicts the date when the sensor will need replacing for
maintaining the measurement accuracy, based on the pH zero and pH slope on each calibration,
and reference impedance (input-2 impedance) after each calibration. The projected replacement
date is predicted based on these parameters stored upon calibration, and displayed the same as
that of projected maintenance. For details, see the description about the projected maintenance.
Since three parameters (pH zero, pH slope, and input-2 impedance after calibration) are used for
this projection, the nearest coming day is selected as the projected replacement date from the
extrapolatedlineofthevaluesobtainedbytheleastsquaresmethod.
n PH module (sensor)
With this screen, you can check the module productions number and software revision of the
installed module.
n HOUSING ASSY
With this screen, you can check the module productions number, software revision, and HART
device revision of the housing assembly.
n Read logbook
The FLXA21 has two types of logbook per sensor to store history information on events, such as
changed settings and calibrations.
Byselectingoneofthelogbooksthatyouwishtocheck,youcanretrieveandcheckthis
information. Storage of history information on each event in a logbook or which logbook to use for
storagecanbesetupontheCongurelogbookscreen.Fordetails,seesection4.5.
PH
3