<Appendix 3 For ISC (Inductive Conductivity)>
App.3-11
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
Theconductance(1/R)isproportionaltothespecicconductivityandaconstantfactorthat
is determined by the geometry of the sensor (length divided by surface area of the hole in the
toroid) and the installation of the sensor.
Thereare2toroidsmountedinthe“doughnut”shapedsensor.Theliquidalsoowsthroughthe
secondtoroidandthereforetheliquidturncanbeconsideredasaprimarywindingofthesecond
ring transformer.
Thecurrentintheliquidwillcreateamagneticeldinthesecondtoroid.Theinducedvoltage
beingtheresultofthismagneticeldcanbemeasuredasanoutput.
Theoutputvoltageofthis“receivecoil”isthereforeproportionaltothespecicconductivityofthe
processliquid.
n Changing the settings
If any setting is accidentally changed, values to the right of the relevant arrow in Table 6 are all
initialized.
Table 6 Parameters that initialize other values
Measurement-> Output: Process
parameter->
Linear: 0% value, 100% value
Table
Communication:HART:PV
Displaysetup:Maindisplay
TrendGraphScreen-> Y-axis(low,high)
Communication: HART
Conguresensor:
Measuuringunit->
Displaysetup:Maindisplay:unit