n Matrix temperature compensation
Table5showsthedefaultsforthematrixtemperaturecompensationselectableinTemperature
compensation (Section 4.2.3).
Table 5 Defaults for the matrix temperature compensation (Reference temperature (Tref.): 25.0°C)
Temp. range Solution 1
(Min.)
Solution 2 Solution 3 Solution 4 Solution 5
(Max.)
Tref. (25.0°C) 6.40 pH 7.00 pH 7.30 pH 7.60 pH 9.00 pH
Tmin. (T1) 5.0°C 6.42 pH 7.38 pH 7.94 pH 8.31 pH 9.74 pH
T2 25.0°C 6.40 pH 7.00 pH 7.30 pH 7.60 pH 9.00 pH
T3 45.0°C 6.34 pH 6.70 pH 6.86 pH 7.06 pH 8.40 pH
T4 65.0°C 6.23 pH 6.45 pH 6.54 pH 6.67 pH 7.91 pH
Tmax. (T5) 85.0°C 6.11 pH 6.25 pH 6.31 pH 6.40 pH 7.51 pH
l User setting
DeletethematrixandthenenternewvaluesasshowninTable6.Grayareasaremandatory.
Other areas can be omitted.
Table 6 Example of user setting (Reference temperature (Tref.): 25.0°C)
Temp. range Solution 1
(Min.)
Solution 2 Solution 3 Solution 4 Solution 5
(Max.)
Tref. (25.0°C) 6.40 pH 7.00 pH 7.30 pH 7.60 pH 9.00 pH
Tmin. (T1) 5.0°C 6.42 pH 7.38 pH 7.94 pH 8.31 pH 9.74 pH
T2 25.0°C 6.40 pH 7.00 pH 7.30 pH 7.60 pH 9.00 pH
T3 45.0°C 6.34 pH 6.70 pH 6.86 pH 7.06 pH 8.40 pH
T4 65.0°C 6.23 pH 6.45 pH 6.54 pH 6.67 pH 7.91 pH
Tmax. (T5) 85.0°C 6.11 pH 6.25 pH 6.31 pH 6.40 pH 7.51 pH
Note: Thegrayareasmustbeentered.
The reference temperature is set in the Temperature Compensation menu (default: 25.0°C).
Input temperatures for compensation in the Temp. ranges menu.
Input values from Solution 1 (low pH) to Solution 5 (high pH).
After entering all values, select “Check values?” to verify that there is no error. This function
checksifthematrixisconsistentlyincrementalordecremental.Ifanyerrorisfound,itslocationis
specied.
Whenthereisnoerror,thematrixiscompensatedlinearlyandtheblanksofthetable(ifany)are
lled.Ifsomeareasareleftemptywithoutrunning“Checkvalues?”,a1st/2ndcomp.matrixerror
will be issued.
n Changing the settings
If any setting is accidentally changed, values to the right of the relevant arrow in Table 7 are all
initialized.
Table 7 Parameters that initialize other values
Sensortype-> Measurement-> Output: Process
parameter->
Linear: 0% value, 100% value
Table
Communication:HART:PV
Displaysetup:Individualdisplay(Maindisplay)
TrendGraphScreen-> Y-axis(low,high)
Communication: HART
Impedance settings