1.4 Regulatory Compliance
Safety: EN61010-1
UL61010-1
CAN/CSAC22.2No.61010-1
EMC:EN61326-1ClassA,Table2(Foruseinindustriallocations)
EN61326-2-3
AS/NZSCISPR11
Korea Electromagnetic Conformity Standard
Installation altitude: 2000 m or less
CategorybasedonIEC61010:I(Note1)
PollutiondegreebasedonIEC61010:2(Note2)
Note1: Installationcategory,calledover-voltagecategory,speciesimpulsewithstandvoltage.
Equipmentwith“CategoryI”(ex.two-wiretransmitter)isusedforconnectiontocircuitsinwhichmeasuresaretaken
to limit transient over-voltages to an appropriately low level.
Note2: Pollutiondegreeindicatesthedegreeofexistenceofsolid,liquid,gasorotherinclusionswhichmayreduce
dielectricstrength.Degree2isthenormalindoorenvironment.
Explosion-proof(Intrinsicallysafetypeandnon-incendive)(forsufxcode:-EA):
ATEX Intrinsically safe approval
Applicable standard
ExplosiveAtmospheres
EN60079-0:2009 Generalrequirements
EN60079-11:2007 Equipmentprotectionbyintrinsicsafety“i”
EN60079-26:2007 Equipmentwithequipmentprotectionlevel(EPL)Ga
EN60529:1992 Degreesofprotectionprovidedbyenclosures(IPCode)
Type of protection
II1GExiaIICGa
Group: II
Category: 1G
T4: for ambient temperature:–20 to 55ºC
T6: for ambient temperature:–20 to 40ºC
Atmosphere pressure: 80 kPa (0.8 bar) to 110 kPa (1.1 bar)
DegreeofProtectionoftheEnclosure:IP66
IECExIntrinsicallysafe
Applicable standard
IEC60079-0:2007 Part0:Generalrequirements
IEC60079-11:2006 Part11:Equipmentprotectionbyintrinsicsafety“i”
IEC 60079-26: 2006 Part 26: Construction, test and marking of Group II Zone 0
electrical apparatus
IEC60529:2001 Degreesofprotectionprovidedbyenclosures(IPCode)
Type of protection
ExiaIICGa
T4: for ambient temperature:–20 to 55ºC
T6: for ambient temperature:–20 to 40ºC
Atmosphere pressure: 80 kPa (0.8 bar) to 110 kPa (1.1 bar)
DegreeofProtectionoftheEnclosure:IP66
FM Intrinsically safe and nonincentive approval
Applicable standard
FM-3600:2011 ApprovalStandardforElectricEquipmentforuseinHazardous
(Classied)LocationsGeneralRequirement
FM-3610: 2010 Approval Standard for Intrinsically Safe Apparatus and Associated
ApparatusforUseinClassI,II,andIII,Division1,Hazardous
(Classied)Locations
FM-3611:2004 NonincendiveElectricalEquipmentforUseinClassIandII,Division2
andClassIII,Divisions1and2,Hazardous(Classied)Locations
FM-3810:2005 ElectricalEquipmentforMeasurement,ControlandLaboratoryUse
NEMA250:1991EnclosuresforElectricalEquipment(1000VoltsMaximum)
ANSI/IEC60529:2004 Degreesofprotectionprovidedbyenclosures(IPCode)
ANSI/ISA60079-02009 Part0:GeneralRequirements
ANSI/ISA60079-112011 Part11:Equipmentprotectionbyintrinsicsafety“i”
1
PH
SC
ISC
DO
SENCOM