User Manual
<19. TROUBLESHOOTING>
19-1
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
19. TROUBLESHOOTING
When a or indicator is displayed on the Main display, detailed error information can
be given by pressing this indicator-button.
See “n Information button ” on page 1-7.
Forsettingoferroritems,refertotheErrorcongurationofthesection4.4,7.3,10.3,13.4,or
16.4.Foreacherroritem,WarningorFaultwillbeindicated.Excepttheseerroritemssetonthe
Errorconguration,someerrorswillbeindicated.
n Installation of sensor module
Whenasensormoduleisnotxedproperly,anerrormessageonthegure19.1maybe
displayed. Check the sensor module installation and lock mechanism.
Ona2-sensormeasurement,ifthedifferenttypesensors(forexample,pHandSC)areinstalled,
anerrormessageonthegure19.2willbedisplayed.
No sensor module
4mA 20mA
Figure 19.1 Error message when no sensor module is installed
Different sensor modules
4mA 20mA
Figure 19.2 Error message for wrong combination of 2 sensor modules
PH
SC
ISC
DO
SENCOM