<16. COMMISSIONING OF SENCOM pH/ORP>
16-10
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
l DeneSENCOMstatus
SettheparametersforanalyzingtheSENCOMsensor.Theyaredisplayedonthedetailscreen.
The parameters are Sterilized temp., Sterilized time, High temp.1, High temp.2, Low pH value,
and High pH value.
16.3 Output setup
Thegeneralprocedureistorstdenethefunctionoftheoutput,OutputorSimulate.Then,set
the process parameters associated with the output. On the Output, an output of measured value
is selected. On the Simulate, a simulation value can be set.
And,theparametersforHOLDfunctioncanbesetonthissetting.
l Output
Theoutputsignalisacurrentvaluespeciedbythefollowingparameters.
Process parameter
The available process parameters depend on the selected “Sensor type” on the Sensor setup
and the selected “Measurement” type on the Measurement setup.
The output of the selected process parameter is shown as a bar on the bottom of the Main
display.Anditsparametersymbol(forexample,PH1)isshownabovethebar,too.When
a selected process parameter is displayed as a measurement value, the top left number or
characteristurnedtobewhitenumberorcharacteronblackbackground(forexample,
). (Refer
to the section 1.2.)
Process parameters can be selected from among:
pH1,Temperature1,andORP1: SENCOMmodule’smeasuredvalues
Setup
Select one of the output methods: Linear and Table.
Linear: Set the 0% and 100% values.
Table: Thisallowsthecongurationofanoutputcurveby21points(5%intervals).
(The 0% and 100% values must be entered.)
Burn
Select the designated output in case of a fault from among Off, Low, and High. See “16.4 Error
conguration”tosettheoutput.
Off: Output depends on the measured value.
Low: Outputisxedto3.6mA
High:Outputisxedto22.0mA.
Damping time
Thisisthetimetakenforaresponsetoastepinputchangetoreach90%ofthenalvalue
(attenuation time). Set this time in sec.
l Simulate
Whenthisfunctionisselected,anoutputoftheinstrumentwillbeaxedcurrentvaluesetin%of
the output span. The output span range is -2.5% to 112.5% (3.6 mA to 22.0 mA).
When “Simulate” is selected, regardless of hold setting, the output is always simulated value.