User Manual

<16. COMMISSIONING OF SENCOM pH/ORP>
16-9
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
16.2.5 Impedance settings
This screen is used to set the impedance relating to an input impedance check.
“High”or“Low”isautomaticallysetaccordingtotheconnectedSENCOMsensor.Youcannot
select the setting. In the case of “Low”, high and low limits can be set.
ForFU20Fsensors,Impedance1is“High”andImpedance2is“Low”.
Input 1 impedance represents the “glass membrane impedance” of a pH sensor. In case of an
ORP sensor, it represents “metal electrode impedance.” Input 2 impedance stands for “reference
impedance.”
The FLXA21 has an impedance check, which is capable of monitoring the impedance of various
sensor systems.
The system can be set to measure the impedances of glass (input 1 impedance: high) and
reference (input 2 impedance: low) electrodes. In applications that tend to leave deposits on the
electrodesortoclogthereferencesensorjunction,theimpedancecheck(seterrorconguration)
on the reference sensor can be used to initiate an alarm, or to initiate the wash cleaning process,
ifoneofthelimitsisexceeded.
16.2.6 Concentration
Generally pH values are not converted into concentrations. Following linear conversion is not
suitable for a lot of measurements.
The FLXA21 is capable of converting a measured pH value into concentration for display.
Parametersrequiredforconcentrationcalculationcanbearbitrarilysetbytheuser.Those
parameters are as follows:
Unit: Thedefaultvalueis%.Theunitcanbeselectedfromamongmg/L,g/L,and
ppm.
pH: The default values are 0 and 14 pH (0 to 14). They are the pH zero and span
requiredforconversiontoconcentration.
Concentration: The default value are 0 and 100%. Concentration for each pH can be set.
Concentration can be assigned to the pH value zero and span individually.
16.2.7 Sensor diagnostic settings
This screen is used to set items relating to sensor diagnostics displayed on the screens invoked
by pressing
.
Gauges are displayed for only parameters that have been enabled in “Sensor diag. settings.”
ParameterssettoDisableareprovidedwithabardisplay.
The setting parameters include Input 1 imp., Input 2 imp., Progress time, and Heat cycle.
Wheninputimpedanceisset“High”(section16.2.5),“FINE”valuecanbechanged.Wheninput
impedance is set “Low”, its “High limit” and “Low limit” will be the limits for diagnostic.
ForFU20Fsensors,Impedance1is“High”andImpedance2is“Low”.
Itisalsopossibletosetthe“Badlimits”oftheprogresstimeandheatcycleandthe“Heatcycle
temp” and “Heat cycle time” of the heat cycle.
16
SENCOM