<16. COMMISSIONING OF SENCOM pH/ORP>
16-8
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
Time
Calibration
Zero, Slope
a
b
c
d
a: Actual change
b: Zero and slope calculated from
the calibration data
c: Zero and slope calculated from
the calibration data and shifted
using the latest calibration data
d: Data from calibration
Figure 16.3 Auto correct
n
ORP settings
l Limits and timing
Zero High/Low
Zero(aspot)highandlowlimits.Duringcalibration,itischeckedwhetherthenewzeroexceeds
thesehighandlowlimits.Narrowingthebandwillpreventbadcalibrationproceduresand
calibration of bad sensors, resulting in higher accuracy. The default values should be adjusted to
suit the application and the “users” criterion.
Slope High/Low
Slope(sensitivity)highandlowlimits.Duringcalibration,itischeckedwhetherthenewslope
exceedsthesehighandlowlimits.Narrowingthebandwillpreventbadcalibrationprocedures
and calibration of bad sensors, resulting in higher accuracy. The default values should be
adjusted to suit the application and the “users” criterion.
Step Range
Set the range over which the stability of a measured value is checked. If variations of a measured
value over the stabilization time are within this setpoint, the measured value is judged to have
stabilized.
Stabilization time
This is the time over which the stability of an ORP value is monitored during calibration. When
variations of the ORP value are within the value set in Step Range over this stabilization time
set here, the value is regarded as being stable. If the value has not stabilized within 10 minutes,
calibration is aborted.
Calibr. interval
Settheintervalinwhichanewcalibrationmusttakeplace.Iftheintervalsethereisexceeded,
theinstrumentwillgiveawarningorfaultaccordingtothesettingontheErrorconguration
screen.
l Zero/Slope
You can enter zero and slope values directly.