<16. COMMISSIONING OF SENCOM pH/ORP>
16-3
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
16.1 Sensor setup
“Sensor type” setup is determined by the sensor to be connected to the instrument. Select one of
the following three sensor types:
pH: Only pH is measured.
ORP: OnlyORPismeasured.UsethissettingfortheORPmeasurementonthebasis
of a glass electrode. For the measurement on the basis of a reference electrode,
selectpH+ORPandthenselectORPin[Measurement].
pH+ORP: BothpHandORParemeasuredsimultaneously.ThissetupalsoallowsrHtobe
measured. ORP measurement is based on a reference electrode
Table 16.2 Selecting sensor types
Sensor type Measurement (Sec. 16.2.1) Description
pH + ORP
pH
ORP For the ORP measurement on the basis of
a reference electrode
pH + ORP
pH + rH
rH
pH (not displayed) Same of pH of “pH + ORP”
ORP (not displayed)
For the ORP measurement on the basis of
a glass electrode
NOTE
Thedefaultof[Sensortype]ispH+ORP.Changethissettingforyoursensors.Notallsensor
typesareavailableforsomeSENCOMsensors.Forexample,pHcannotbeselectedwhenan
ORP sensor is connected.
NOTE
The selection of “Sensor type” determines the menu structure throughout the instrument.
WhenpH+ORPisspeciedasthesensortype,theobjectundermeasurementwillbedisplayed
at the top of “Measurement setup.”
IfthesensortypeischangedfrompHtoORP,partoftherelevantsetupisreset,requiring
reconguration.
WhenORPisspecied,andwhenanORPsensordoesn’thaveanytemperatureelement,
setbothof“Temperaturetoohigh/low”in“Errorconguration”toOffsothatnorelatederroris
caused. (Refer to the section 16.4.) Without temperature input, a temperature shown on the Main
displayisthemaximumortheminimumtemperatureofthetemperatureelement’srangeseton
theTemperaturesettings.Settingin“Displaysetup”canmaketemperaturevalueonthedisplay
not to be displayed. (Refer to the section 16.7.)
Setting the Safe sensor disconnection to “Enable” allows operators to disconnect sensors. The
defaultis“Disable.”
Whenthissettingis“Disable,”thesafesensordisconnectionisnotguaranteedbecausesensors
are performing measurement or automatic update parameters are being updated.
When this setting is “Enable,” sensors are not performing measurement or automatic update
parameters are not being updated, thus sensors can be disconnected safely.