User Manual
<16. COMMISSIONING OF SENCOM pH/ORP>
16-1
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
16.
COMMISSIONING OF SENCOM pH/ORP
This chapter describes how to check and change settings from the Commissioning screen.
When you move to the Commissioning screen, the output is held.
Execute:
Calibration/Wash
HOLD
Setup:
Comissioning
Change language
Start Quick Setup
Execute &Setup
Sensor setup
Measurement setup
Output setup
Error configuration
Logbook configuration
Advanced setup
Display setup
Commissioning
Figure 16.1 Commissioning screen
Operations in Commissioning can be password-protected. If you set up a password, always take
a note of it. For details on setting a password, see section 16.6.3.
Figure15.1showsthecommissioningprocedure.Beforechanginganyparameters,readthe
relevant sections in this document and understand how the change of parameters affects the
performance of this instrument. If you set a wrong value, return it to the default setting or value
and then set it again.
Ontherststartup,theparametersarealldefaultvaluesconguredatthefactory.Checkthe
parameters in Table 16.1 and change any of them if necessary depending on the sensors to be
connected and the purpose of use.
Set“Sensortype”rst.Somemeasurementparametersandrelevantoptionschange
accordingly.TheunderlinedparametersinTable16.1areusedforthequicksetup.
Thedefaultvaluesandsettingrangesarelistedinthe“UsersettingtableofSENCOM”inthe
attachedCD-ROM.
Afterconrmingthattheinstrumentoperatesnormallywiththeparameters,printoutthe“User
settingtableofSENCOM”andwritedowntheseparametersinthecolumnofUserSettings.
All user parameters can also be saved in the instrument.
SelectCommissioning→Advancedsetup→Settings→Saveusersettings(seeSection
16.6.1).
16
SENCOM