<13. COMMISSIONING OF DO (Dissolved Oxygen)>
13-1
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
13. COMMISSIONING OF DO (Dissolved
Oxygen)
This chapter describes how to check and change settings from the Commissioning screen.
When you move to the Commissioning screen, the output is held.
Execute:
Calibration/Wash
HOLD
Temporary output
Setup:
Comissioning
Change language
Start Quick Setup
Execute &Setup
Sensor setup
Measurement setup
Output setup
Error configuration
Logbook configuration
Advanced setup
Display setup
Calculated data setup
Commissioning
Figure 13.1 Example of the commissioning screen (for two modules)
Operations in Commissioning can be password-protected. If you set up a password, always take
a note of it. For details on setting a password, see section 13.6.3.
Figure12.1showsthecommissioningprocedure.Beforechanginganyparameters,readthe
relevant sections in this document and understand how the change of parameters affects the
performance of this instrument. If you set a wrong value, return it to the default setting or value
and then set it again.
Ontherststartup,theparametersarealldefaultvaluesconguredatthefactory.Checkthe
parameters in Table 13.1 and change any of them if necessary depending on the sensors to be
connected and the purpose of use.
Set“Sensortype”rst.Somemeasurementparametersandrelevantoptionschange
accordingly.TheunderlinedparametersinTable13.1areusedforthequicksetup.
Thedefaultvaluesandsettingrangesarelistedinthe“UsersettingtableofDO”intheattached
CD-ROM.
Afterconrmingthattheinstrumentoperatesnormallywiththeparameters,printoutthe“User
settingtableofDO”andwritedowntheseparametersinthecolumnofUserSettings.
All user parameters can also be saved in the instrument.
SelectCommissioning→Advancedsetup→Settings→Saveusersettings(seeSection
13.6.1).