<10. COMMISSIONING OF ISC (Inductive Conductivity)>
10-1
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
10. COMMISSIONING OF ISC (Inductive
Conductivity)
This chapter describes how to check and change settings from the Commissioning screen.
When you move to the Commissioning screen, the output is held.
Execute:
Calibration
HOLD
Setup:
Comissioning
Change language
Start Quick Setup
Execute &Setup
Measurement setup
Output setup
Error configuration
Logbook configuration
Advanced setup
Display setup
Commissioning
Figure 10.1 Commissioning screen
Operations in Commissioning can be password-protected. If you set up a password, always take
a note of it. For details on setting a password, see section 10.5.3.
Figure9.1showsthecommissioningprocedure.Beforechanginganyparameters,readthe
relevant sections in this document and understand how the change of parameters affects the
performance of this instrument. If you set a wrong value, return it to the default setting or value
and then set it again.
Ontherststartup,theparametersarealldefaultvaluesconguredatthefactory.Checkthe
parameters in Table 10.1 and change any of them if necessary depending on the sensors to be
connected and the purpose of use.
Set“Measurement”rst.Somemeasurementparametersandrelevantoptionschange
accordingly.TheunderlinedparametersinTable10.1areusedforthequicksetup.
Thedefaultvaluesandsettingrangesarelistedinthe“UsersettingtableofISC”intheattached
CD-ROM.
Afterconrmingthattheinstrumentoperatesnormallywiththeparameters,printoutthe“User
settingtableofISC”andwritedowntheseparametersinthecolumnofUserSettings.
All user parameters can also be saved in the instrument.
SelectCommissioning→Advancedsetup→Settings→Saveusersettings(seeSection
10.5.1).