User Manual

<9. OPERATION OF ISC (Induvtive Conductivity)>
9-9
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
n Read logbook
The FLXA21 has two types of logbook per sensor to store history information on events, such as
changed settings and calibrations.
Byselectingoneofthelogbooksthatyouwishtocheck,youcanretrieveandcheckthis
information. Storage of history information on each event in a logbook or which logbook to use for
storagecanbesetupontheCongurelogbookscreen.Fordetails,seesection10.4.
History information on events are automatically stored on the preset conditions.
Inadditiontothisstoring,followingthreemessagescanbemanuallystoredinthelogbook;
"Sensor washed by hand", "Module replaced", "Sensor replaced"
To store these messages, press
and select one of three messages from the Item on the
Memorandum screen. Its event date/time will be the time when a message is selected and
entered.
WhenapasswordforCommissioningissetonthepasswords’setup,pressing
requires
entering the password. (Refer to the section 10.5.3.)
Read Logbook:
logbook1-1
Memorandum:
Item
Sensor washed by hand
Enter?
No
logbook1-1 1/1
Memorandum:
Item
Sensor replaced
Enter?
No
2010/02/15 17:04
Sensor replaced
2010/02/15 15:15
Power on
Finish
2010/02/15 15:15
Power on
Finish
logbook1-1 1/1
** When storing message manually.
**
(Fig.9.5)
Sensor washed by hand
Module replaced
Sensor replaced
logbook1-1
logbook1-2
No
Yes
Figure 9.7 Detail display (continued)
9.5 Trend graphics
Pressing on the Zoom display changes the display to a graphical mode in which the
average measured value is shown on a time scale. The “Live” value is also digitally displayed in
atextbox.Thetimescale(X-axis)andtheprimaryvaluescale(Y-axis)aresetinthe“DISPLAY
SETUP”menu(Section10.7.2).
ISC
9