User Manual

<7. COMMISSIONING OF SC (Conductivity)>
7-14
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
l Main display
When one module is installed on the instrument, only the Main display is available.
Three measurement values can be set to display on the Main display as a primary value (1st
line), a second value (2nd line) and a third value (3rd line) respectively.
Onthe“Additionaltext”,atextofupto12alphanumericcharacterscanbeassignedtoeach
measurement value.
AdditionaltextsaredisplayedontheMaindisplay,andareusefulforidentifyingmeasurements.
Insomecases,notall12characterscanbedisplayedduetotheletters;checkthetexts
displayedontheMaindisplayaftersetting.Ifapartofthetextismissing,adjustthenumberof
characters.
l Dual display
This screen is used to set items to be displayed in the top part (1st row) and bottom part (2nd
row) of the Home display. These settings are available when two sensors are installed in the
instrument.
When the “Empty” is selected for the 2nd row, the Home display can not be displayed.
l Individual display
When two modules are installed on the instrument, display items on the Main display for each
module can be set on these settings.
On the “SC1 display” or “SC2 display” screen, three measurement values can be set to display
on each Main display as a primary value (1st line), a second value (2nd line) and a third value
(3rd line) respectively.
Onthe“Additionaltext”,atextofupto12alphanumericcharacterscanbeassignedtoeach
measurement value.
AdditionaltextsaredisplayedontheMaindisplay,andareusefulforidentifyingmeasurements.
Insomecases,notall12characterscanbedisplayedduetotheletters;checkthetexts
displayedontheMaindisplayaftersetting.Ifapartofthetextismissing,adjustthenumberof
characters.
7.6.2 Trend
This screen is used to make settings for the Trend Graph Screen.
Set the process parameters to be displayed for each trend. They can be set for the 1st to 3rd
trends. When all three process parameters are set “Empty”, there is no trend display (no trend
button).
l X-axis: Timing
SelecttheX-axistiming’stimespanonthetrendgraphdisplayfromalist.
l Y-axis: Limits
SettheY-axishighandlowlimitsonthetrendgraphdisplayonaTrendscreenbasis.
NOTE