User Manual

<7. COMMISSIONING OF SC (Conductivity)>
7-9
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
DuringtheCommissioningortheQuickSetup,themAoutputisautomaticallyheld.Thepreset
valuedependsonasettingonthe“Lastorxed”.
“Last”: The preset value is a value measured just before hold condition.
“Fixed”: Thepresetvalueisavaluesetinthe“FixedvaluemA”.
Whenthe“Fixed”isselected,setamAvalueinthe“FixedvaluemA”.
Selection on the “Hold during Calibration” decides to activate or deactivate the hold function
automatically during calibration.
“Enabled”: Activation of the automatic hold function
“Disabled”: Noautomaticholdfunction
7.3 Errorconguration
InErrorconguration,congurethestatusesofvariouserrorcauses.
This allows the system to notify the user of the occurrence of an error according to the status
categoriesintheErrorconguration.
Select a status category from among Off, Warn. (Warning), and Fault.
“Fault”automaticallyperformsburn-out.WhenBurnhasbeensettoOff(7.2Outputsetup),only
the error message is displayed.
“Warn.” displays an error message.
When selecting PH201G in the communication setting, make sure that the “Fail contact” setting
is appropriate.
The settable causes of errors are determined based on the settings of the Sensor setup and
Measurement setup, and a status category is set to the causes displayed in the Errors 1/3 to 3/3
screens.
Table7.3 Errorconguration
Display item Description Default
Conductivity too high
(or Concentration)
Conductivity or resistivity is lower than the minimum limit. Warn.
Conductivity too low
(or Concentration)
Conductivityorresistivityishigherthanthemaximumlimit. Warn.
Temperature too high Measuredprocesstemperatureishigherthanthemaximumlimit. Warn.
Temperature too low Measured process temperature is lower than the minimum limit. Warn.
Polarization detec Polarization of the electrodes is detected. Off
Calibr.timeexceeded Calibrationtimeexceedsthecalibrationinterval(seeSection7.1.5). Off
USP ConductivityexceedstheUSPlimit. Off
Congurationerror:
1stcomp.matrix
2ndcomp.matrix
Thetemperaturecompensationmatrixisnotdenedproperly(see
Section 7.1.4).
Fault
Concentration table Theconcentrationtableisnotdenedproperly(seeSection7.1.6). Fault
CAUTION
Ifcancelinganerrorcongurationcouldberisky,donotcancelitasadangeroussituationmay
result.
SC
7