3. Commissioning
QUICK SET-UP NOTES:
a. pH measurement module
Under“measurement“aselectionismadeforpH,ORPorpH&ORP.Theselected
measurement must be in accordance with the sensor wiring. When rH measurement is
requestedpH&ORPmustbechosenonthislevel.TherHmustthenbeselectedinthe
commissioning menu.
b. SC measurement module
Under“measurement”aselectionismadebetweenConductivity,Resistivity,Concentration
or Concentration plus Conductivity. On this level only Conductivity or Resistivity can be
selected. Settings for Concentration measurement must be done in the commissioning
menu.
c. DO measurement module
Under“sensortype”aselectionismadeforGalvanicorPolarographic.Theselectedsensor
type must be in accordance with the sensor wiring. Otherwise the analyzer or sensor can be
damaged.