Specifications
YCO-1047-03-13
PPC-200(W) Operation Manual
The optional analog meter has three programmable over and under
registers labelled A, B and C. They are selected with the ,
and keys on the meter face. Pressing one of these
keys once will cause the LED display to ash the under and over
setpoints for that key. The meter must be calibrated before it can be
set. Both procedures are detailed below.
Calibrating the Analog Meter
1) Press and hold the key 2-3 seconds. The display will
read Adj for adjust and the indicator will move to the under range.
2) Use the and keys to align the indicator with the
low end of the under range.
3) Press the key. The indicator will move to the over range.
4) Use the and keys to align the indicator with the
high end of the over range.
5) Press the key to save the calibration and to return to
normal weighing mode.
V Optional Analog Meter
* * * WARNING * * *
The PPC-200W is NOT WASHDOWN if the analog meter has been installed! Damage will result to the
scale and meter if an attempt is made to washdown the scale after it has been modied for the analog
meter.
NOTICE
The analog meter option is NOT LEGAL FOR TRADE.
ACCEPT
UNDER
OVER
ACCEPT
UNDER
OVER
ACCEPT
UNDER
OVER
8