Manual

45
Operation Method
Calibration Offset Function
3
Set calibration offset
TROUBLE
ENTER
CANCEL
Calibrate +2.5
(Ex.)
Indicated temperature on controller: 37
Measured temperature: 36
Calibration offset: -1
Press the ENTER key. The setting is decided
and the function menu selection screen is
displayed.
Offset is adjustable within the range from +3.0 to
-3.0.
Temperature correction in
overheating prevention device:
The temperature of overheating prevention device is
automatically corrected by the same amount when
that of controller is collected.